Configuring the RF Signal
R&S
®
SMW200A
574User Manual 1175.6632.02 ─ 30
User correction cannot be activated simultaneously with absolute level compensation
based on user-defined frequency response values. These functions exclude each
other; you can one only of them at a time.
See user manual R&S SMW-K544 User-Defined Frequency Response Correction.
Remote command:
[:SOURce<hw>]:CORRection[:STATe] on page 1119
User Correction
Indicates the corrected level value for a specific frequency point.
Remote command:
[:SOURce<hw>]:CORRection:VALue? on page 1118
UCOR Data ...
Accesses the standard "File Select" function of the instrument. The provided navigation
possibilities in the dialog are self-explanatory.
Files with user correction values are files with predefined file extension *.uco. When a
file is selected, the dialog indicates the filename.
You can create the file internally in the table editor or externally.
●
To select an existing file, select "Select List > navigate to the file *.uco > Select"
●
Use the general editor function to create internally new file or to edit an existing
one.
●
Use the standard file manager function to load externally created files to the instru-
ment.
Remote command:
[:SOURce]:CORRection:CSET:CATalog? on page 1119
[:SOURce<hw>]:CORRection:CSET[:SELect] on page 1118
[:SOURce]:CORRection:CSET:DELete on page 1119
Edit UCOR Data ...
Opens the build-in table editor to define a new correction table or edit an existing one.
See also:
●
Chapter 8.7.7, "List Editor", on page 524
●
"Fill..." on page 526
●
Chapter 8.9.3.3, "Fill with Sensor", on page 578
8.9.3.2 List Editor
The "User Correction" and "List Mode" dialogs provide a build-in list editor for defining
the corresponding value pairs.
The table and navigation functions of the UCOR and list mode editors have the same
structure and are therefore summarized below. The access is specified for both modes,
the fields and functions are explained using the example of the list mode.
Improving Level Performance