Configuring the RF Signal
R&S
®
SMW200A
595User Manual 1175.6632.02 ─ 30
2. Select "UCOR Data" > "UCOR Data...."
3. Create a file: "New" > Filename and confirm with "OK".
4. Select the new file with "Select".
5. Select "Edit UCOR Data...".
6. In the "Edit User Correction Data A ucor", select "Edit" > "Fill".
The "Fill Table" dialog enables you to fill in the values of the columns automatically.
7. To fill in the frequency column:
a) Select "Select Column To Fill > Frequency / Hz".
b) Select "Range > e.g. 15" to determine the number of values.
c) Select "Start Value > e.g. 1.56 GHz".
d) Select "Increment Value > e.g. 1 MHz", to determine the frequency steps.
e) Select "Fill", to insert the frequency values.
The "Fill" function also fills the column of the "Correction Value / dB" values with a
predefined value, since empty cells lead to the data loss of the entire line. If you
need different values, you can change them manually, or you can use the auto-
matic function "Fill with Sensor" as described in the next steps.
8. Select "Fill with Sensor".
The "Fill User Correction Data With Sensor" dialog provides an overview of the
sensor configuration.
9. Select "Execute".
Improving Level Performance