Multi-Instrument Setups
R&S
®
SMW200A
643User Manual 1175.6632.02 ─ 30
"No Sync"
Synchronization is not achieved, because:
The baseband generators in the secondary instruments or in the
primary instruments are not activated, see "To start signal genera-
tion" on page 648.
There is no physical connection (the synchronization signal is not
connected as required).
For details, see Chapter 10.3.1.2, "How to Connect and Configure
Instruments in Primary-Secondary Instrument Mode",
on page 644.
The secondary instrument does not detect the synchronization
signal.
Remote command:
:SCONfiguration:MULTiinstrument:TRIGger:SYNChronization?
on page 914
Global Connector Settings
Option: R&S SMW-B10 or R&S SMW-B9 with DACW board with revision < 5.
Provides a quick access to the related global connector settings.
With "State = On", the synchronization signal is assigned per default as follows:
"Multi Instrument Trigger > Primary"
"Connector = USER 6": "Direction = Output", "Signal = Baseband Sync Out"
"Multi Instrument Trigger > Secondary"
"Connector = USER 6": "Direction = Output", "Signal = Baseband Sync Out"
"Connector = USER 5": "Direction = Input", "Signal = Baseband Sync In"
See also:
Chapter 10.3.1.2, "How to Connect and Configure Instruments in Primary-Secon-
dary Instrument Mode", on page 644.
See also Chapter 12.2, "Configuring Local and Global Connectors", on page 714.
Show Connector Baseband Sync In/Out
Option: R&S SMW-B9 with DACW board with revision 6.
Displays the physical location of the connectors on the rear panel of the instrument.
The connectors are fixed and cannot be changed. The following connectors are used:
Sync Out connector: ADV TRIG
Sync In connector: ADV CLK.
See also:
Chapter 10.3.1.2, "How to Connect and Configure Instruments in Primary-Secon-
dary Instrument Mode", on page 644
"Connect the instruments and distribute the synchronization signal (R&S SMW-B9,
DACW revision 6)" on page 646.