Multi-Instrument Setups
R&S
®
SMW200A
657User Manual 1175.6632.02 ─ 30
Remote command:
:SCONfiguration:RFALignment:SETup:INFO:CALibration:PARameters?
on page 940
Time since Last Calibration
Indicates the elapsed time since the moment the calibration described in the loaded
setup file is performed.
Remote command:
:SCONfiguration:RFALignment:SETup:INFO:CALibration:AGE?
on page 940
Temperature Offset to Last Calibration
Indicates the difference in temperature since the moment the calibration described in
the loaded setup file is performed.
The indication is color coded; if the temperature deviation exceeds +/- 3K, the value is
displayed in red.
Remote command:
:SCONfiguration:RFALignment:SETup:INFO:CALibration:TEMPerature:
OFFSet? on page 940
Frequency/PEP Control
Comprises the frequency and PEP-related settings.
The displayed values resemble the values you have set during the calibration process.
Mode ← Frequency/PEP Control
Indicates the method used to define the calibrated frequency and PEP ranges.
"Range"
By the min and max values of the range.
"List"
As a sequence of discrete frequency values.
See Frequency Values/PEP Values.
Consider that greater number of calibration points leads to longer cal-
ibration time.
Remote command:
:SCONfiguration:RFALignment:SETup:INFO:CALibration:FREQuency:
MODE? on page 940
:SCONfiguration:RFALignment:SETup:INFO:CALibration:POWer:MODE?
on page 940
Frequency Range/PEP Range ← Frequency/PEP Control
Indicates the frequency and PEP ranges the calibrated values apply for.
This value is displayed if "Mode = Range" is used during the calibration, where the
range is defined by the min and max values.
Remote command:
:SCONfiguration:RFALignment:SETup:INFO:CALibration:FREQuency:
RANGe:LOWer? on page 941
:SCONfiguration:RFALignment:SETup:INFO:CALibration:FREQuency:
RANGe:UPPer? on page 941
Generating Phase-Coherent Synchronous and Aligned Signals