Multi-Instrument Setups
R&S
®
SMW200A
659User Manual 1175.6632.02 ─ 30
Current Corrections Settings
Access:
► Select "RF Ports Alignment > Info" > "Current Corrections".
Figure 10-7: RF Ports Setup Info > Current Corrections
This dialog displays summary information on the current correction values per RF
path.
Settings:
Calibrated Frequency..................................................................................................659
Calibrated Level (PEP)................................................................................................659
Baseband Frequency Offset....................................................................................... 660
Digital Attenuation.......................................................................................................660
Correction Values........................................................................................................660
└ Phase............................................................................................................660
└ Level............................................................................................................. 660
└ I/Q Delay.......................................................................................................660
Frequency Response Correction................................................................................ 660
Calibrated Frequency
Indicates the frequency f
Cal
the calibration data is valid for.
This is the value you have set as RF frequency during the calibration process.
Remote command:
:SOURce<hw>:RFALignment:CALibrated:FREQuency? on page 1203
Calibrated Level (PEP)
Indicates the level P
Cal
the calibration data is valid for.
Generating Phase-Coherent Synchronous and Aligned Signals