File and Data Management
R&S
®
SMW200A
675User Manual 1175.6632.02 ─ 30
To reset an individual parameter: Open its context-sensitive menu and select "Pre-
set this parameter...".
"Factory Preset"
A factory preset is the most profound preset function that resets almost all instru-
ment settings, including reference oscillator, network and remote access settings.
Retained are the following settings:
Security, password, and settings protected by these passwords
User-defined data, like setups or data lists
Settings that relate to an integration of the instrument in a measurement setup.
To restore the factory defaults, select System Config > Setup > Settings > Factory
Preset.
Note: Perform a "Factory Preset" only if it is necessary. After a "Factory Preset",
the network connection to the instrument no longer exists.
Presetting the instrument to a user-defined instrument state
The reset functions set the parameters and operating modes to default values prede-
fined by the factory. Alternatively to these default settings, you can:
Define user-specific recall settings to be restored after a preset
(see Chapter 11.2.3, "How to Recall User Settings Automatically after Preset",
on page 677)
Store and reload user-defined instrument states
(see Chapter 11.4.2, "How to Save and Recall Instrument Settings", on page 684)
Mark / Do not mark parameters changed from preset
To survey the current state of the settings concerning default values, the R&S SMW
offers a feature that visually identifies deviations from the default values.
For more information, see Chapter 11.2.2, "How to Identify Parameters Which Are Not
in a Preset State", on page 676.
11.2.1 Preset, Set to Default and Factory Preset Settings
Preset..........................................................................................................................675
Set To Default..............................................................................................................676
Preset this Parameter................................................................................................. 676
Execute Factory Preset...............................................................................................676
Preset
Resets all parameters and switching states, and closes all opened dialogs.
Restoring the (Default) Instrument Configuration