General Instrument Functions
R&S
®
SMW200A
731User Manual 1175.6632.02 ─ 30
Threshold Clock/Trigger Input
Sets the high/low threshold in volts for the trigger and clock signal inputs of the base-
band section.
Remote command:
[:SOURce<hw>]:INPut:TM:CLOCk:LEVel on page 952
[:SOURce<hw>]:INPut:TM:TRIGger:LEVel on page 952
Impedance Clock/Trigger Input
Selects the input impedance for the external trigger/clock inputs.
Remote command:
[:SOURce<hw>]:INPut:TM:CLOCk:IMPedance on page 953
[:SOURce<hw>]:INPut:TM:TRIGger:IMPedance on page 953
Clock Input Slope
Sets the polarity of the active slope of an externally applied clock signal.
Custom digital modulation with external serial data (Data Source > "External"): A
higher bit modulation extern serial data requires a reference symbol clock signal to
generate a bit clock out of it.
●
This parameter sets the active edge of the symbol clock.
●
To set the active edge of the bit clock, use the parameter Bit Clock Slope.
●
Observe the impact of this parameter on the preview graphic in the "Baseband >
Custom Digital Mod > Data Source" dialog.
See also "External Modulation Data" on page 226.
Remote command:
[:SOURce<hw>]:INPut:TM:CLOCk:SLOPe on page 953
Trigger Input Slope
Sets the polarity of the active slope of an externally applied trigger signal.
Remote command:
[:SOURce<hw>]:INPut:TM:TRIGger:SLOPe on page 954
12.2.7 How to Enable Signals and Perform Signal to Connector Mapping
This section introduces the general principle of connector configuration.
General Workflow
The Figure 12-2 shows the main configuration stages, together with the corresponding
user interface dialogs.
Configuring Local and Global Connectors