General Instrument Functions
R&S
®
SMW200A
756User Manual 1175.6632.02 ─ 30
"Confirm Unsigned"
Performs the signature verification.
If the check detects any discrepancies, the instrument issues a warn-
ing message. You can still update the firmware or reject updating.
This setting also enables you to downgrade the firmware version.
"All Packages"
Accepts all packages without signature verification.
"R&S Signed Packages"
Performs the signature check.
If the check detects any discrepancies, the instrument issues a warn-
ing message and locks the update to this firmware.
Remote command:
:SYSTem:SECurity:SUPolicy on page 1235
12.6.2.2 Disk & Memory Security Settings
Access:
► Select "System Config > Setup > Security > Security > General > Disk & Memory".
The "Disk & Memory" tab secures controlled access to the mass memory and pre-
vents information from leaving a secure environment.
The remote commands available to control security settings are described in Chap-
ter 14.20, "SYSTem Subsystem", on page 1217.
USB Storage
Activates the access to external USB storage media.
To apply the change: enter the security password and confirm with "Accept". Otherwise
the change has no effect.
See also Chapter 11.9.4, "Using a USB Storage Device for File Transfer", on page 701.
Note: Remove all USB memory devices before disabling the USB storage. If any USB
memory device remains connected, disabling is blocked, and the instrument returns a
warning message.