Remote Control Commands
R&S
®
SMW200A
843User Manual 1175.6632.02 ─ 30
The figures in Table 14-1 highlight the differences. The figures illustrate the same
LxMxN configuration but address the entities in different way.
Table 14-1: Addressing entities in remote control
SOURce<hw> only ENTity<ch>:SOURce<hw> combination
LxMxN is the short form of the used system configuration,
where:
●
L represents the Entity
●
M the Baseband
●
N the Stream
*) applies for:
●
:SOUR:BB:FOFF|POFF|PGA
●
:SOUR:BBIN:...
●
:SOUR:BB:..., if separate baseband sources are used
Example:
●
"System Configuration > Mode > Advanced"
●
"Entities = 2", "Basebands = 2", "Streams = 2"
●
"BB Source Config > Separate Source"
The command :ENTity2:SOURce1:BB:EUTRa:STATe 1 enables the generation of
an EUTRA/LTE signal in the first baseband within the second entity.
It is equivalent to :SOURce3:BB:EUTRa:STATe 1.
Logically, also the following command pairs are equivalent to each other:
●
:ENTity2:SOURce1:FSIMulator:STATe
and :SOURce2:FSIMulator:STATe
●
:ENTity2:SOURce1:AWGN:STATe and :SOURce3:AWGN:STATe
●
:ENTity2:SOURce1:BB:FOFF and :SOURce3:BB:FOFF
●
:ENTity2:SOURce1:BBIN:DIG:SOUR and :SOURce3:BBIN:DIG:SOUR
You can find the corresponding commands also with the "Show SCPI command" func-
tions in the context menu of parameters.
See "Findig out the SCPI command using "Show SCPI Command"" on page 828.