Remote Control Commands
R&S
®
SMW200A
847User Manual 1175.6632.02 ─ 30
Determines whether the contents of the ENABle registers are preserved or reset when
the instrument is switched on. Thus a service request can be triggered when the instru-
ment is switched on, if the status registers ESE and SRE are suitably configured. The
query reads out the contents of the "power-on-status-clear" flag.
Parameters:
<Action> 0 | 1
0
The contents of the status registers are preserved.
1
Resets the status registers.
*RCL <Number>
Recall
Loads the instrument settings from an intermediate memory identified by the specified
number. The instrument settings can be stored to this memory using the command
*SAV with the associated number.
It also activates the instrument settings which are stored in a file and loaded using the
MMEMory:LOAD <number>, <file_name.extension> command.
Manual operation:
See "Recall Immediate x" on page 683
*RST
Reset
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
The command is equivalent to SYSTem:PRESet.
Usage:
Setting only
Manual operation: See "Preset" on page 675
*SAV <Number>
Save
Stores the current instrument settings under the specified number in an intermediate
memory. The settings can be recalled using the command *RCL with the associated
number.
To transfer the stored instrument settings in a file, use the command :MMEMory:
STORe:STATe.
Manual operation:
See "Save Immediate x" on page 683