Remote Control Commands
R&S
®
SMW200A
861User Manual 1175.6632.02 ─ 30
Usage: Query only
:MMEMory:DELete <Filename>
Removes a file from the specified directory.
Setting parameters:
<Filename> string
String parameter to specify the name and directory of the file to
be removed.
Example: See Example "Working with files and directories" on page 854.
Usage: Event
SCPI confirmed
Manual operation: See "Cut, Copy&Paste and Delete" on page 692
:MMEMory:LOAD:STATe
<SavRclStateNumb>, <file_name>
Loads the specified file stored under the specified name in an internal memory.
After the file has been loaded, the instrument setting must be activated using an *RCL
command.
Setting parameters:
<SavRclStateNumb> Determines to the specific <number> to be used with the *RCL
command, e.g. *RCL 4.
<file_name> String parameter to specify the file name with extension
*.savrcltxt.
Example: See Example "Saving and loading current settings"
on page 853.
Usage: Setting only
Manual operation: See "Recall" on page 683
:MMEMory:MDIRectory <Directory>
Creates a subdirectory for mass memory storage in the specified directory. If no direc-
tory is specified, a subdirectory is created in the default directory. This command can
also be used to create a directory tree.
Setting parameters:
<Directory> string
String parameter to specify the new directory.
Example: See Example "Working with files and directories" on page 854.
Usage: Event
Manual operation: See "Create New Directory" on page 692