Remote Control Commands
R&S
®
SMW200A
864User Manual 1175.6632.02 ─ 30
:CALibration<hw>:ALL:TIME?......................................................................................... 865
:CALibration:ALL:WARN................................................................................................. 865
:CALibration:DATA:FACTory:DATE?.................................................................................865
:CALibration<hw>:CONTinueonerror................................................................................866
:CALibration<hw>:DEBug............................................................................................... 866
:CALibration<hw>:IQModulator:LOCal?............................................................................866
:CALibration<hw>:FREQuency:CONVerter:EXTernal?....................................................... 867
:CALibration<hw>:LOSCillator:COUPling:LOCal?..............................................................867
:CALibration:ALL[:MEASure]? [<Force>]
Starts all internal adjustments that do not need external measuring equipment.
Note: If an external frequency converter R&S SZU is connected, the internal adjust-
ments are blocked, and the R&S SMW generates an error message.
Query parameters:
<Force> string
Return values:
<Measure> 0 | 1 | OFF | ON
Example:
CAL:ALL:MEAS?
// 0
// Executes the adjustments of all instrument functions.
// When completed, it indicates that the adjustment
// has been performed successfully.
Usage: Query only
Manual operation: See "Adjust All" on page 1280
:CALibration<hw>:ALL:DATE?
Queries the date of the most recently executed full adjustment.
Return values:
<Date> string
Example:
CAL:ALL:DATE?
// "2016-01-01"
Usage: Query only
Manual operation: See "Last Full Adjustment" on page 1280
:CALibration<hw>:ALL:INFormation?
Queries the current state of the internal adjustment.
Return values:
<CalInfoText> string
CALibration Subsystem