Remote Control Commands
R&S
®
SMW200A
866User Manual 1175.6632.02 ─ 30
Example:
CAL:DATA:FACT:DATE?
// "2016-01-01"
Usage: Query only
Manual operation: See "Last Factory Calibration" on page 1264
:CALibration<hw>:CONTinueonerror <State>
Continues the calibration even though an error was detected. By default adjustments
are aborted on error.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 0)
Example:
CAL:CONT ON
// Continues calibration after an error
Manual operation: See "Continue Adjustment on Error" on page 1282
:CALibration<hw>:DEBug <State>
Activates logging of the internal adjustments.
Setting parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
CALibration:DEBug 1
Usage: Setting only
Manual operation: See "Log Debug Info" on page 1282
:CALibration<hw>:IQModulator:LOCal?
Starts adjustment of the I/Q modulator for the currently set frequency and baseband
gain. The I/Q modulator is adjusted with respect to carrier leakage, I/Q imbalance and
quadrature.
Return values:
<Local> 0 | 1 | OFF | ON
*RST: 0
Example:
// Start adjustment for the I/Q modulator
// for the currently set frequency
CALibration:IQModulator:LOCal?
// 0
// Adjustments are performed successfully
Usage: Query only
Manual operation: See "Adjust I/Q Modulator Current Frequency" on page 451
CALibration Subsystem