Remote Control Commands
R&S
®
SMW200A
883User Manual 1175.6632.02 ─ 30
:HCOPy:FILE[:NAME]:AUTO:DIRectory:CLEar
Deletes all files with extensions *.bmp, *.jpg, *.png and *.xpm in the directory set
for automatic naming.
Example: See Example "Store a hard copy of the display" on page 879
Usage: Event
Manual operation: See "Clear Path" on page 705
:HCOPy:FILE[:NAME]:AUTO:FILE?
Queries the name of the automatically named hard copy file.
An automatically generated file name consists of:
<Prefix><YYYY><MM><DD><Number>.<Format>.
You can activate each component separately, to individually design the file name.
Return values:
<File> string
Example: See Example "Store a hard copy of the display" on page 879.
Usage: Query only
:HCOPy:FILE[:NAME]:AUTO:STATe <State>
Activates automatic naming of the hard copy files.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 1
Example: See Example "Store a hard copy of the display" on page 879
Manual operation: See "Automatic Naming" on page 703
:HCOPy:FILE[:NAME]:AUTO[:FILE]:DAY:STATe <State>
:HCOPy:FILE[:NAME]:AUTO[:FILE]:MONTh:STATe <State>
:HCOPy:FILE[:NAME]:AUTO[:FILE]:YEAR:STATe <State>
Uses the date parameters (year, month or day) for the automatic naming. You can acti-
vate each of the date parameters separately.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 1
Example: See Example "Store a hard copy of the display" on page 879
Manual operation: See "Prefix, Year, Month, Day" on page 705
HCOPy Subsystem