Remote Control Commands
R&S
®
SMW200A
886User Manual 1175.6632.02 ─ 30
:OUTPut<hw>[:STATe] <State>
Activates the RF output signal.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
:OUTPut1 ON
Activates the RF output.
Manual operation: See "RF State/RF ON" on page 460
:OUTPut<hw>[:STATe]:PON <Pon>
Defines the state of the RF output signal when the instrument is switched on.
Parameters:
<Pon> OFF | UNCHanged
*RST: n.a. (factory preset: UNCHanged)
Example:
:OUTPut1:PON OFF
The RF output is deactivated when the instrument is switched
on.
Manual operation: See "Power-On State" on page 713
:OUTPut<hw>:AMODe <AMode>
Sets the step attenuator mode at the RF output.
Note: The setting [:SOURce<hw>]:POWer:ATTenuation:RFOFf:MODE
FATTenuation has higher priority than the attenuator modes FIXed and MANual.
Parameters:
<AMode> FIXed | MANual | AUTO
AUTO
The step attenuator adjusts the level settings automatically,
within the full variation range.
FIXed
The step attenuator and amplifier stages are fixed at the current
position, providing level settings with constant output VSWR.
The resulting variation range is calculated according to the posi-
tion.
To use this mode, activate the ALC (see [:SOURce<hw>]:
POWer:ALC[:STATe]).
MANual
You can set the level manually, in 10 dB steps.
*RST:
AUTO
Example:
:SOURce1:POWer:ALC:STATe 1
:OUTPut1:AMODe FIXed
OUTPut Subsystem