Remote Control Commands
R&S
®
SMW200A
887User Manual 1175.6632.02 ─ 30
Manual operation: See "Mode" on page 564
:OUTPut<hw>:BLANk:LIST:STATe <State>
Activates RF output blanking.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 1
Example:
:OUTPut1:BLAN:LIST:STAT ON
Activates the RF output blanking.
:OUTPut<hw>:IMPedance?
Queries the impedance of the RF outputs.
Return values:
<Impedance> G1K | G50 | G10K
*RST: G50
Example:
:OUTPut1:IMP?
queries the impedance of RF output.
Response: 50
the impedance is 50 ohms
Usage: Query only
Manual operation: See "RF output impedance" on page 461
:OUTPut<hw>:AFIXed:RANGe:LOWer?
:OUTPut<hw>:AFIXed:RANGe:UPPer?
Queries the settable minimum/maximum value in mode :OUTPut:AMODe FIXed, i.e.
when the attenuator is not being adjusted.
See :OUTPut<hw>:AMODe on page 886
Return values:
<Upper> float
Increment: 0.01
Default unit: dBm
Example:
:OUTPut1:AMODe FIXed
:OUTPut1:AFIXed:RANGe:UPPer?
// -27
:OUTPut1:AFIXed:RANGe:LOW?
// -50
Usage: Query only
Manual operation: See "Level Range" on page 469
OUTPut Subsystem