Remote Control Commands
R&S
®
SMW200A
894User Manual 1175.6632.02 ─ 30
:SLISt:CLEar[:ALL]
Removes all R&S NRP power sensors from the list.
Example: See Example "Detecting and assigning a power sensor"
on page 890.
Usage: Event
Manual operation: See "Clear" on page 585
:SLISt:ELEMent<ch>:MAPPing
<Mapping>
Assigns an entry from the :SLISt[:LIST]? to one of the four sensor channels.
Parameters:
<Mapping> SENS1 | SENSor1 | SENS2 | SENSor2 | SENS3 | SENSor3 |
SENS4 | SENSor4 | UNMapped
Sensor channel.
*RST: UNMapped
Example: See Example "Detecting and assigning a power sensor"
on page 890.
Manual operation: See "Sensor Mapping List" on page 584
:SLISt:SENSor:MAP <SensorId>, <Mapping>
Assigns a sensor directly to one of the sensor channels, using the sensor name and
serial number.
To find out the the sensor name and ID, you can get it from the label of the R&S NRP,
or using the command :SLISt:SCAN[:STATe]. This command detects all R&S NRP
power sensors connected in the LAN or via 'USBTMC protocol.
Setting parameters:
<SensorId> string
<Mapping> enum
Example: See Example "Detecting and assigning a power sensor"
on page 890.
Usage: Setting only
Manual operation: See "Sensor Mapping List" on page 584
:INITiate<hw>[:POWer]:CONTinuous <Continuous>
Switches the local state of the continuous power measurement by R&S NRP power
sensors on and off. Switching off local state enhances the measurement performance
during remote control.
SENSe, READ, INITiate and SLISt Subsystems