Remote Control Commands
R&S
®
SMW200A
897User Manual 1175.6632.02 ─ 30
Manual operation: See "S-Parameter" on page 593
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe <State>
Activates the use of the S-parameter correction data.
Note: If you use power sensors with attenuator, the instrument automatically activates
the use of S-parameter data.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
:SENSe1:POWer:CORRection:SPDevice:STATe 1
Activates the use of the S-parameters correction data.
Manual operation: See "S-Parameter" on page 593
:SENSe<ch>[:POWer]:CORRection:SPDevice:LIST?
Queries the list of the S-parameter data sets that have been loaded to the power sen-
sor.
Return values:
<List> string list
*RST: 0
Usage: Query only
Manual operation: See "S-Parameter" on page 593
:SENSe<ch>[:POWer]:DISPlay:PERManent:PRIority <Priority>
Selects average or peak power for permanent display.
Parameters:
<Priority> AVERage | PEAK
*RST: AVERage
Example:
:SENS1:DISP:PERM:STAT ON
Turns on the permanent view.
:SENS1:DISP:PERM:PRI AVER
Sets the average power for display.
Manual operation: See "Display" on page 590
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe <State>
Activates the permanent display of the measured power level results. The instrument
also indicates the sensor type, the connection, the measurement source and the offset
if set.
SENSe, READ, INITiate and SLISt Subsystems