Remote Control Commands
R&S
®
SMW200A
903User Manual 1175.6632.02 ─ 30
:SENSe<ch>[:POWer]:TYPE?
Queries the sensor type. The type is automatically detected.
Return values:
<Type> string
Example:
:SENS1:TYPE?
Queries the type of sensor.
Response: NRP-Z21
The R&S NRP-Z21 sensor is used.
Usage: Query only
Manual operation: See "Sensor type and serial number" on page 589
:SENSe<ch>[:POWer]:ZERO
Performs zeroing of the sensor.
Zeroing is required after warm-up, i.e. after connecting the sensor.
Note: Switch off or disconnect the RF power source from the sensor before zeroing.
We recommend that you zero in regular intervals (at least once a day), if:
The temperature has varied more than about 5 °C.
The sensor has been replaced.
You want to measure very low power.
Example:
:SENS1:ZERO
Executes zeroing.
Usage: Event
Manual operation: See "Zero" on page 590
14.18 SCONfiguration Subsystem
The SCONfiguration subsystem contains the commands for defining the system
configuration setting, like stream mapping and used baseband configuration.
Required options
See "Required options" on page 124.
Suffixes
The following suffixes are used: