Remote Control Commands
R&S
®
SMW200A
913User Manual 1175.6632.02 ─ 30
// save the configuration
*SAV 1
:MMEMory:STORe:STATe 1,"/var/user/8x8_MIMO_Subset1.savrcl.txt"
// transfer to file to the second instrument
// load the configuration and change the subset
SCONfiguration:MIMO:SUBSet SET2
// trigger the baseband signal generation and hence the fading process
OUTPut1:USER6:TRIGger:IMMediate
● Multi Instrument Configuration.............................................................................. 913
● Baseband and Fading Configuration.................................................................... 914
● I/Q Stream Mapping..............................................................................................919
● External RF and I/Q Instruments...........................................................................921
● RF Ports Alignment...............................................................................................935
14.18.1 Multi Instrument Configuration
Option: R&S SMW-B9.
The following commands configure multi-instrument settings.
Commands:
:SCONfiguration:MULTiinstrument:MODE.........................................................................913
:SCONfiguration:MULTiinstrument:STATe......................................................................... 913
:SCONfiguration:MULTiinstrument:TRIGger:SYNChronization?...........................................914
:SCONfiguration:MULTiinstrument:MODE <MSMode>
Sets if the instrument works as a primary or as a secondary instrument.
Parameters:
<MSMode> PRIMary | SECondary
*RST: PRIMary
Example: See Example "Configuring primary-secondary mode"
on page 907.
Manual operation: See "Multi Instrument Trigger" on page 642
:SCONfiguration:MULTiinstrument:STATe <TriggerState>
Activates the selected mode.
Parameters:
<TriggerState> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Configuring primary-secondary mode"
on page 907.