Remote Control Commands
R&S
®
SMW200A
926User Manual 1175.6632.02 ─ 30
:SCONfiguration:EXTernal:CODer<ch>:IQConnection:STATe?
:SCONfiguration:EXTernal:FADer<ch>:IQConnection:STATe?
:SCONfiguration:EXTernal:BBMM<ch>:IQConnection:STATe?
Queries the status of the I/Q connection of the digital interfaces.
Return values:
<IQConnState> 0 | 1 | OFF | ON
Example: See Example "Connecting and configuring external instrument"
on page 905.
Usage: Query only
Manual operation: See "I/Q Connection" on page 140
See "RF State" on page 143
:SCONfiguration:EXTernal:CODer<ch>:RCONnection:STATe?
:SCONfiguration:EXTernal:FADer<ch>:RCONnection:STATe?
:SCONfiguration:EXTernal:BBMM<ch>:RCONnection:STATe?
:SCONfiguration:EXTernal:RF<ch>:RCONnection:STATe?
:SCONfiguration:EXTernal:IQOutput<ch>:RCONnection:STATe?
Queries the status of the remote connection.
Return values:
<State> 0 | 1 | OFF | ON
Example: See Example "Connecting and configuring external instrument"
on page 905.
Usage: Query only
Manual operation: See "Remote Connection Status" on page 141
:SCONfiguration:EXTernal:CODer<ch>:INAMe?
:SCONfiguration:EXTernal:FADer<ch>:INAMe?
:SCONfiguration:EXTernal:BBMM<ch>:INAMe?
:SCONfiguration:EXTernal:RF<ch>:INAMe?
:SCONfiguration:EXTernal:IQOutput<ch>:INAMe?
Queries the name of the connected external instrument.
Return values:
<InstrName> string
Returns the name of the connected external instrument.
<InstrmentName> (SerialNumber) <Path>
the instrument name, as retrieved via the DIG I/Q interface
<InstrmentName>[, <RfPath>] or <InstrmentName> (Serial-
Number)
the instrument name, as defined in with the "Remote Config"
settings or as defined by the command :SCONfiguration:
EXTernal:REMote:ADD
SCONfiguration Subsystem