Remote Control Commands
R&S
®
SMW200A
938User Manual 1175.6632.02 ─ 30
Manual operation: See "Setup" on page 652
:SCONfiguration:RFALignment:SETup:CATalog?
Queries the names of the existing setup files in the default directory.
Per default, the instrument saves user-defined files in the /var/user/ directory. Use
the command :MMEM:CDIRectory to change the default directory to the currently
used one.
Only files with extension *.rfsa are listed.
Example: See Example "Loading and activating setup files" on page 935.
Usage: Query only
Manual operation: See "Setup" on page 652
:SCONfiguration:RFALignment:SETup:FILE <SetupFile>
Loads the selected file from the default or the specified directory. Loaded are files with
extension *.rfsa.
Parameters:
<SetupFile> "<filename>"
Filename or complete file path; file extension can be omitted.
Query the filenames of existing setup files with the command :
SCONfiguration:RFALignment:SETup:CATalog?.
Example: See Example "Loading and activating setup files" on page 935.
Manual operation: See "Setup" on page 652
:SCONfiguration:RFALignment:SETup:STATus?
Queries information on the internal compensation status and the connected secondary
instruments.
Return values:
<SetupStatus> NALign | ALIGned | ERRor | WARNing | INACtive | NOSetup |
INValid
NOSetup = Setup is not loaded
INValid = Loaded setup does not match the current setup
INACtive = Setup loaded but RF port alignment not enabled
NALign = Setup is loaded but aligned not triggered
ALIGned | ERRor | WARNing = Setup is aligned, error,
warning
*RST: n.a. (factory preset)
Example: See Example "Loading and activating setup files" on page 935.
Usage: Query only
SCONfiguration Subsystem