Remote Control Commands
R&S
®
SMW200A
939User Manual 1175.6632.02 ─ 30
Manual operation: See "Status" on page 654
:SCONfiguration:RFALignment:ALIGn
If a valid setup is loaded and RF ports alignment is enabled, sends the multi instrument
trigger signal to all instruments to synchonize the basebands of the instruments.
Example: See Example "Loading and activating setup files" on page 935.
Usage: Event
Manual operation: See "Align" on page 654
:SCONfiguration:RFALignment:STATe
<State>
If a valid setup file is selected, applies the specified correction data.
Load the setup file with the command :SCONfiguration:RFALignment:SETup:
FILE.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 0)
Example: See Example "Loading and activating setup files" on page 935.
Manual operation: See "State" on page 652
:SCONfiguration:RFALignment:SETup:INFO:CALibration:DATE?
Queries the time the calibration described in the loaded setup file is performed.
Return values:
<Date> string
Example: See Example "Retrieving details on the calibrated values"
on page 935.
Usage: Query only
Manual operation: See "Calibration Date/Time" on page 656
:SCONfiguration:RFALignment:SETup:INFO:CALibration:TIME?
the date and time the calibration described in the loaded setup file is performed.
Return values:
<Time> string
Example: See Example "Retrieving details on the calibrated values"
on page 935.
Usage: Query only
Manual operation: See "Calibration Date/Time" on page 656
SCONfiguration Subsystem