Remote Control Commands
R&S
®
SMW200A
941User Manual 1175.6632.02 ─ 30
RANGe
Range, see:
:SCONfiguration:RFALignment:SETup:INFO:
CALibration:FREQuency:RANGe:LOWer? on page 941
:SCONfiguration:RFALignment:SETup:INFO:
CALibration:FREQuency:RANGe:UPPer? on page 941
:SCONfiguration:RFALignment:SETup:INFO:
CALibration:FREQuency:STEP? on page 942
:SCONfiguration:RFALignment:SETup:INFO:
CALibration:POWer:RANGe:LOWer? on page 942
:SCONfiguration:RFALignment:SETup:INFO:
CALibration:POWer:RANGe:UPPer? on page 942
:SCONfiguration:RFALignment:SETup:INFO:
CALibration:POWer:STEP? on page 942
LIST
List, see:
:SCONfiguration:RFALignment:SETup:INFO:
CALibration:FREQuency:LIST:VALues? on page 941
:SCONfiguration:RFALignment:SETup:INFO:
CALibration:POWer:LIST:VALues? on page 941
Example:
See Example "Retrieving details on the calibrated values"
on page 935.
Usage: Query only
Manual operation: See "Mode" on page 657
:SCONfiguration:RFALignment:SETup:INFO:CALibration:FREQuency:LIST:
VALues?
:SCONfiguration:RFALignment:SETup:INFO:CALibration:POWer:LIST:VALues?
Return values:
<ListOfValues> string
Example:
SCONfiguration:RFALignment:SETup:INFO:CALibration:LOCoupling?
// NACTive
SCONfiguration:RFALignment:SETup:INFO:CALibration:POWer:MODE?
// LIST
SCONfiguration:RFALignment:SETup:INFO:CALibration:POWer:LIST:VALues?
// -21,-15,-10
Usage: Query only
Manual operation: See "Frequency Values/PEP Values" on page 658
:SCONfiguration:RFALignment:SETup:INFO:CALibration:FREQuency:RANGe:
LOWer?
:SCONfiguration:RFALignment:SETup:INFO:CALibration:FREQuency:RANGe:
UPPer?
SCONfiguration Subsystem