Remote Control Commands
R&S
®
SMW200A
949User Manual 1175.6632.02 ─ 30
MARKA1|MARKC1|MARKA2|MARKC2|(MARKA3|MARKC3) =
Baseband A/C Marker 1/2/(3)
SCLA|SCLB|SCLC|SCLD = Symbol Clock A/B/C/D
Option: R&S SMW-B10
LATTA|LATTB|LATTC|LATTD = Lev Att A/B/C/D
BGATA|BGATB|BGATC|BGATD = Burst Gate A/B/C/D
HOPA|HOPB|HOPC|HOPD = HOP A/B/C/D
CWMODA|CWMODB|CWMODC|CWMODD = CW/Mod A/B/C/D
TRIGA|TRIGB|TRIGC|TRIGD = Triggered A
The character A/B/C/D in the parameter value indicates the
baseband the signal is related to.
*RST:
depends on TM suffix
Example:
:SOURce1:INPut:TM2:DIRection OUTPut
:OUTPut1:TM2:DIRection?
// OUTPut
:OUTPut1:TM2:SIGNal MARKA2
Options: TMC4/5/6, LATTA|LATTB|LATTC|LATTD| BGATA|BGATB|
BGATC|BGATD| HOPA|HOPB|HOPC|HOPD| CWMODA|
CWMODB|CWMODC|CWMODD| TRIGA|TRIGB|TRIGC|TRIGD
require R&S SMW-B10
Manual operation: See "Signal" on page 729
[:SOURce]:INPut:USER<ch>:SIGNal <Signal>
Determines the control signal that is input at the selected connector.
To define the connector direction, use the command [:SOURce]:INPut:USER<ch>:
DIRection.
Suffix:
USER<ch>
.
1 to 6
Parameters:
<Signal> TRIG1 | TRIG2 | CLOCK1 | CLOCK2 | NSEGM1 | NSEGM2 |
NONE | FEEDback | IPULSA | IPULSB | ERRTA | ERRTB |
BERDATIN | BERCLKIN | BERDATENIN | BERRESTIN |
SYNCIN
TRIG1|TRIG2 = Global Trigger 1/2
CLOCK1|CLOCK2 = Global Clock 1/2
NSEGM1|NSEGM2 = Global Next Segment 1/2
IPULSA|IPULSB = Pulse In A/B, available for USER4|5|6
FEEDback = Baseband Feedback, available for USER6
SYNCIN = Baseband Sync In
BERDATIN|BERCLKIN|BERDATENIN|BERRESTIN = BER
Data, Clock, Data Enable and Restart
ERRTA|ERRTB = External restart trigger signals for REG
SOURce Subsystem