Remote Control Commands
R&S
®
SMW200A
951User Manual 1175.6632.02 ─ 30
BERRESTOUT|BERDATENOUT|BERCLKOUT|BERDATOUT
require R&S SMW-K80
RTRIGA|RTRIGB require R&S SMW-K78
Manual operation: See "Signal" on page 725
:OUTPut:USER<ch>:TRIGger[:IMMediate]
Generates a short pulse signal and outputs it at the USER connector.
This signal can serve as a common external trigger signal for triggering of several
R&S SMW, see Example "Triggering several R&S SMW instruments simultaneously"
on page 649.
Suffix:
USER<ch>
.
6
Example: See Example "Global connectors settings configuration"
on page 945.
Usage: Event
Manual operation: See "Execute Trigger" on page 727
[:SOURce]:INPut:USER:CLOCk:LEVel <Level>
[:SOURce]:INPut:USER:TRIGger:LEVel <Level>
Sets the threshold for any input signal at the USER1-3 connectors.
Parameters:
<Level> float
Range: 0.1 to 2
Increment: 0.1
*RST: 1
Example: See Example "Global connectors settings configuration"
on page 945.
Manual operation: See "Threshold USER1-3 Input" on page 728
[:SOURce]:INPut:USER:PULM:LEVel <Level>
Sets the threshold for any input signal at the USER4-6 connectors.
Parameters:
<Level> float
Range: 0.1 to 2
Increment: 0.1
*RST: 1
Default unit: V
Example: See Example "Global connectors settings configuration"
on page 945.
SOURce Subsystem