Remote Control Commands
R&S
®
SMW200A
952User Manual 1175.6632.02 ─ 30
Manual operation: See "Threshold USER4-6 Input/Threshold Pulse Input"
on page 542
[:SOURce]:INPut:USER:CLOCk:IMPedance <Impedance>
[:SOURce]:INPut:USER:TRIGger:IMPedance <Impedance>
Selects the input impedance for the external trigger inputs.
Parameters:
<Impedance> G1K | G50
*RST: G1K
Example: See Example "Global connectors settings configuration"
on page 945.
Manual operation: See "Impedance Clock/Trigger Input" on page 728
[:SOURce]:INPut:USER:CLOCk:SLOPe
<Slope>
Sets the polarity of the active slope of an externally applied clock signal.
Parameters:
<Slope> NEGative | POSitive
*RST: POSitive
Example: See Example "Global connectors settings configuration"
on page 945.
Manual operation: See " Clock Input Slope" on page 728
[:SOURce]:INPut:USER:TRIGger:SLOPe <Slope>
Sets the polarity of the active slope of an applied instrument trigger.
Parameters:
<Slope> NEGative | POSitive
*RST: POSitive
Example: See Example "Global connectors settings configuration"
on page 945
Manual operation: See " Trigger Input Slope" on page 728
[:SOURce<hw>]:INPut:TM:CLOCk:LEVel <Level>
[:SOURce<hw>]:INPut:TM:TRIGger:LEVel <Level>
Sets the high/low threshold in volts for the trigger and clock signal inputs of the base-
band section.
SOURce Subsystem