Remote Control Commands
R&S
®
SMW200A
956User Manual 1175.6632.02 ─ 30
// Set the sample rate per channel.
SOURce1:BBIN:SRATe:SOURce USER
SOURce1:BBIN:CHANnel0:SRATe 500000000
SOURce1:BBIN:SRATe:SUM?
// 500000000
SOURce1:BBIN:MPERiod 2
SOURce1:BBIN:ALEVel:EXECute
[:SOURce<hw>]:BBIN:STATe.......................................................................................... 956
[:SOURce<hw>]:BBIN:DIGital:SOURce............................................................................957
[:SOURce<hw>]:BBIN:DIGital:INTerface...........................................................................957
[:SOURce<hw>]:BBIN:CDEVice?.....................................................................................957
[:SOURce<hw>]:BBIN:CHANnel<ch0>:BB:STATe..............................................................957
[:SOURce<hw>]:BBIN:CHANnel<ch0>:NAME...................................................................958
[:SOURce<hw>]:BBIN:CHANnel<ch0>:POWer:CFACtor.................................................... 958
[:SOURce<hw>]:BBIN:CHANnel<ch0>:POWer:PEAK........................................................958
[:SOURce<hw>]:BBIN:CHANnel<ch0>:POWer:RMS......................................................... 959
[:SOURce<hw>]:BBIN:CHANnel<ch0>:SRATe.................................................................. 959
[:SOURce<hw>]:BBIN:IQSWap[:STATe]........................................................................... 959
[:SOURce<hw>]:BBIN:MODE..........................................................................................960
[:SOURce<hw>]:BBIN:SRATe:SOURce............................................................................960
[:SOURce<hw>]:BBIN:SRATe:SUM?................................................................................960
[:SOURce<hw>]:BBIN:SRATe:MAX?................................................................................961
[:SOURce<hw>]:BBIN:SRATe[:ACTual]............................................................................ 961
[:SOURce<hw>]:BBIN:DIGital:ASETting:STATe.................................................................961
[:SOURce<hw>]:BBIN:MPERiod......................................................................................962
[:SOURce<hw>]:BBIN:ALEVel:EXECute...........................................................................962
[:SOURce<hw>]:BBIN:POWer:CFACtor............................................................................962
[:SOURce<hw>]:BBIN:POWer:PEAK............................................................................... 963
[:SOURce<hw>]:BBIN:POWer:RMS?...............................................................................963
[:SOURce<hw>]:BBIN:OLOad:STATe?............................................................................. 963
[:SOURce<hw>]:BBIN:OLOad:HOLD:STATe?................................................................... 963
[:SOURce<hw>]:BBIN:OLOad:HOLD:RESet.....................................................................964
[:SOURce<hw>]:BBIN:STATe <State>
Enables feeding of an external digital signal into the signal path.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Applying an external digital baseband signal (DIG
I/Q)" on page 955.
Manual operation: See "State" on page 161