Remote Control Commands
R&S
®
SMW200A
958User Manual 1175.6632.02 ─ 30
Parameters:
<BbinIqHsChanSta> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Applying an external digital baseband signal (HS
DIG I/Q)" on page 955.
Options: R&S SMW-B9
Manual operation: See "BB" on page 171
[:SOURce<hw>]:BBIN:CHANnel<ch0>:NAME
<BbinIqHsChanNam>
Queries the channel name.
Parameters:
<BbinIqHsChanNam> string
Example: See Example "Applying an external digital baseband signal (HS
DIG I/Q)" on page 955.
Options: R&S SMW-B9
Manual operation: See "Name" on page 170
[:SOURce<hw>]:BBIN:CHANnel<ch0>:POWer:CFACtor <BbinIqHsChCrFac>
Sets the crest factor of the individual channels.
Parameters:
<BbinIqHsChCrFac> float
Range: 0 to 30
Increment: 0.01
*RST: 0
Example: See Example "Applying an external digital baseband signal (HS
DIG I/Q)" on page 955.
Options: R&S SMW-B9
Manual operation: See "Crest Factor" on page 171
[:SOURce<hw>]:BBIN:CHANnel<ch0>:POWer:PEAK <BbinHsChPoPeak>
Sets the peak level per channel.
Parameters:
<BbinHsChPoPeak> float
Range: -60 to 3.02
Increment: 0.01
*RST: 0
Example: See Example "Applying an external digital baseband signal (HS
DIG I/Q)" on page 955.