Remote Control Commands
R&S
®
SMW200A
960User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:BBIN:MODE <Mode>
Defines that a digital external signal is applied.
Parameters:
<Mode> DIGital
*RST: DIGital
Example: See Example "Applying an external digital baseband signal (DIG
I/Q)" on page 955.
[:SOURce<hw>]:BBIN:SRATe:SOURce
<Source>
Selects whether the sample rate is estimated based on the digital input signal or is a
user-defined value.
Parameters:
<Source> DIN | HSDin | USER
USER
Enabled for [:SOURce<hw>]:BBIN:DIGital:INTerface
DIN.
Set the value with [:SOURce<hw>]:BBIN:SRATe[:ACTual].
DIN
Enabled for [:SOURce<hw>]:BBIN:DIGital:SOURce
CODER1|CODER2.
Estimates the sample rate based on the digital input signal.
HSDin
Enabled for [:SOURce<hw>]:BBIN:DIGital:INTerface
HSDin.
*RST:
HSDin
*RST:
USER (R&S SMW-B10)/HSDin (R&S SMW-B9)
Example:
See Example "Applying an external digital baseband signal (DIG
I/Q)" on page 955.
Options: HSDin requires R&S SMW-B9
Manual operation: See "Sample Rate Source" on page 162
[:SOURce<hw>]:BBIN:SRATe:SUM?
Queries the sum of the sample rates of all active channels.
Return values:
<DigIqHsOutSRSum> integer
Range: 0 to depends on settings
*RST: 0
Example: See Example "Applying an external digital baseband signal (HS
DIG I/Q)" on page 955.
SOURce Subsystem