Remote Control Commands
R&S
®
SMW200A
963User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:BBIN:POWer:PEAK <Peak>
Peak level of the external baseband signal relative to full scale of 0.5 V (in terms of dB
full scale).
Parameters:
<Peak> float
Range: -60 to 3.02
Increment: 0.01
*RST: 0
Default unit: dBfs
Example: See Example "Applying an external digital baseband signal (DIG
I/Q)" on page 955
Manual operation: See "Peak Level" on page 164
[:SOURce<hw>]:BBIN:POWer:RMS?
Queries the RMS level of the external digital baseband signal.
Return values:
<Rms> float
Range: -100 to 10
Increment: 0.01
*RST: 0
Example: See Example "Applying an external digital baseband signal (DIG
I/Q)" on page 955.
Usage: Query only
Manual operation: See "Level" on page 165
[:SOURce<hw>]:BBIN:OLOad:STATe?
Queries the current overflow state.
Return values:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: see Example "Applying an external digital baseband signal (DIG
I/Q)" on page 955
Usage: Query only
Manual operation: See "Signal Monitoring" on page 165
[:SOURce<hw>]:BBIN:OLOad:HOLD:STATe?
Queries an overload since the last reset for evaluating the measurement.