General Commands
[:SOURce<hw>]:BB:DM:STATe <State>
Enables/disables digital modulation. Switching on digital modulation turns off all the
other digital standards in the same signal path.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Performing general tasks" on page 989
Manual operation: See "State" on page 264
[:SOURce<hw>]:BB:DM:PRESet
Sets the default settings for digital modulation (*RST values specified for the com-
mands).
Not affected is the state set with the command SOURce<hw>:BB:DM:STATe
Example:
See Example "Performing general tasks" on page 989
Usage: Event
Manual operation: See "Set To Default" on page 264
[:SOURce<hw>]:BB:DM:SRATe <SRate>
Sets the symbol rate in Hz/kHz/MHz or sym/s, ksym/s and Msym/s.
Parameters:
<SRate> float
Range: 50 to depends on the installed options
Increment: 1E-3
*RST: 270833.333
Default unit: Hz or sym/s
Example: See Example "Performing general tasks" on page 989
Manual operation: See "Symbol Rate" on page 265
[:SOURce<hw>]:BB:DM:STANdard <Standard>
Selects predefined set of settings according to the selected standard, see Table 5-12.