Command reference
R&S
®
ZNA
1101User Manual 1178.6462.02 ─ 29
Example:
*RST; SWE:TYPE POW
CALC:PAR:SDEF 'Trc1', 'a1'
Reset the instrument, activate a power sweep, and select a
wave quantity a
1
for the trace Trc1.
DISP:WIND:TRAC:FEED 'Trc1'
Display the generated trace in the active window.
CALC:MATH:SDEF 'StimVal'; STAT ON
Define a mathematical trace and display it instead of the active
data trace.
CALC:MATH:WUN ON
Take into account that the stimulus value is a voltage (derived
from the source power) rather than a dimensionless quantity.
The y-axis range of the mathematical trace now exactly corre-
sponds to the power sweep range.
Manual operation:
See "Result is Wave Quantity" on page 446
7.3.1.11 CALCulate:PARameter...
The CALCulate:PARameter... commands assign names and measurement
parameters to traces. The commands are device-specific.
CALCulate<Ch>:PARameter:CATalog?...........................................................................1101
CALCulate<Ch>:PARameter:CATalog:SENDed?............................................................. 1102
CALCulate<Ch>:PARameter:DEFine:SGRoup................................................................ 1102
CALCulate<Ch>:PARameter:DELete..............................................................................1103
CALCulate:PARameter:DELete:ALL............................................................................... 1104
CALCulate<Ch>:PARameter:DELete:CALL.....................................................................1104
CALCulate<Ch>:PARameter:DELete:CMEMory...............................................................1104
CALCulate:PARameter:DELete:MEMory.........................................................................1105
CALCulate<Ch>:PARameter:DELete:SGRoup................................................................ 1105
CALCulate<Ch>:PARameter:MEASure...........................................................................1105
CALCulate<Ch>:PARameter:MEASure:SENDed............................................................. 1106
CALCulate<Ch>:PARameter:SDEFine............................................................................1106
CALCulate<Ch>:PARameter:SDEFine:SENDed.............................................................. 1110
CALCulate<Ch>:PARameter:SELect...............................................................................1111
CALCulate<Ch>:PARameter:CATalog?
Returns the trace names and measurement parameters of all traces assigned to a par-
ticular channel.
The result is a string containing a comma-separated list of trace names and measure-
ment parameters, e.g. 'CH4TR1,S11,CH4TR2,S12'. The measurement parameters
are returned according to the naming convention of CALCulate<Ch>:PARameter:
SDEFine. The order of traces in the list reflects their creation time: The oldest trace is
the first, the newest trace is the last trace in the list.
Suffix:
<Ch>
.
Channel number. If unspecified the numeric suffix is set to 1.
SCPI command reference