1117User Manual 1178.6462.02 ─ 29
Example: See CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:
STARt
Manual operation: See "Range List" on page 490
CALCulate<Chn>:RIPPle:SEGMent<Seg>:RESult?
Returns the result of the ripple limit check in the previously defined limit range no.
<Seg>. The response consists of two parameters:
<Boolean> 0 for "passed", 1 for "failed".
<Limit> measured ripple in the limit range. A result is returned even if the limit
check in the range no. <Seg> is disabled; see example below.
A reset deletes all ripple limit ranges. Use CALCulate<Ch>:RIPPle:FAIL? to query
the result for global ripple limit check.
Note: In remote control, the ripple limit check result is calculated once at the end of
each sweep. If the ripple limits are changed, a new sweep is required to obtain upda-
ted ripple limit check results. In single sweep mode (INITiate<Ch>:CONTinuous
OFF), the new sweep must be started explicitly. This behavior is different from manual
control where a changed ripple limit line can directly affect the pass/fail result of the
displayed trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg> Number of the ripple limit range.
Example:
*RST; CALC:RIPP:DATA 1, 1500000000, 2000000000,
3
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
CALC:RIPP:STAT ON; SEGM:RES?
Enable the limit check and query the result for the created
range. Possible response: 0,0.3529814004.
CALC:RIPP:DATA 0, 2500000000, 3000000000, 3
Define a second ripple limit range with disabled limit check (no
limit check results are displayed in the diagram area).
CALC:RIPP:SEGM2:RES?
Query the result for the second range. Possible response:
0,1.149071925.
Usage:
Query only
Manual operation: See "Ripple Check" on page 488
CALCulate<Chn>:RIPPle:SEGMent<Seg>[:STATe] <Boolean>
Enables or disables the limit check in the ripple limit range no. <Seg>.
Suffix:
<Chn>
.
Channel number used to identify the active trace