Command reference
R&S
®
ZNA
1122User Manual 1178.6462.02 ─ 29
Parameters:
<Boolean> ON | OFF - statistical info field on or off.
*RST: OFF
Example: See CALCulate<Chn>:STATistics[:STATe]
Manual operation: See "Min/Max/Peak-Peak, Mean/Std Dev/RMS" on page 456
CALCulate<Chn>:STATistics:FORMat <Format>
For complex-valued traces (Smith, Polar) this determines how the MEAN, STDDev,
MAX, MIN, RMS and PTPeak statistics are calculated, see CALCulate<Chn>:
STATistics:RESult?.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Format> ZVAB | IMPedance | ADMittance
ZVAB
The results are based on unformatted wave quantities (voltages)
IMPedance
The results are based on resistance values
ADMittance
The results are based on conductance vaules
Manual operation: See "Format" on page 457
CALCulate<Chn>:STATistics:NLINear:COMP:LEVel <dBValue>
Defines the compression value x for the compression point measurement for dB for-
matted traces.
Use CALCulate<Chn>:STATistics:NLINear:COMP:RESult? to retrieve the
compression results.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
dBValue Compression value
Example: See CALCulate<Chn>:STATistics:NLINear:COMP:
RESult?
Manual operation: See "Compr. Point / Compr. Val." on page 459
CALCulate<Chn>:STATistics:NLINear:COMP:PHASe <PhaseValue>
Defines the compression value x for the compression point measurement for phase
formatted traces.
SCPI command reference