Command reference
R&S
®
ZNA
1224User Manual 1178.6462.02 ─ 29
<SeqNr> Sequence number, defining the order in which the commands
shall be executed.
For every channel, segment and RFFE interface, the sequence
numbers must be consecutive, starting at 1; reusing a sequence
number overwrites a previously defined command.
Parameters:
<Delay> Delay time
Default unit: s
Manual operation: See "Wait (Sweep Sequencer Table)" on page 667
CONTrol<Ch>:SEGMent<SegNr>:SEQuence<SeqNr>:GPIO<Port>:VOLTage
<Voltage>
For segmented sweeps this command allows to define the GPIO voltage settings to be
applied at the start of each segment.
The GPIO ports can be enabled/disabled using CONTrol<Ch>:GPIO<Port>[:
STATe]. Complementary RFFE commands can be defined using CONTrol<Ch>:
SEGMent<SegNr>:SEQuence<SeqNr>:RFFE<Bus>:COMMand:DATA.
Suffix:
<Ch>
.
Channel number
<SegNr> Segment number
<SeqNr> Sequence number, defining the order in which the commands
shall be executed.
For every channel, segment and RFFE interface, the sequence
numbers must be consecutive, starting at 1; reusing a sequence
number overwrites a previously defined command.
Parameters:
<Voltage> Range: -6 to +12 V
Increment: 0.005 V
Default unit: V
Manual operation: See "GPIO columns (sweep sequencer table)" on page 666
CONTrol<Ch>:SEGMent<SegNr>:SEQuence<SeqNr>:RFFE<Bus>:COMMand:
DATA <Command>
For segmented sweeps this command allows to define the RFFE command(s) to be
executed at the start of each segment.
Command execution on an RFFE interface can be enabled/disabled using
CONTrol<Ch>:RFFE<Bus>:SETTings[:STATe]. Complementary GPIO switches
can be defined using CONTrol<Ch>:SEGMent<SegNr>:SEQuence<SeqNr>:
GPIO<Port>:VOLTage.
Suffix:
<Ch>
.
Channel number
SCPI command reference