Command reference
R&S
®
ZNA
1376User Manual 1178.6462.02 ─ 29
Setting parameters:
<TestPort1>,
<TestPort2>
VNA test port numbers
Example: See Chapter 8.2.5.5, "Inline calibration", on page 1754.
Usage: Setting only
[SENSe:]CORRection:COLLect:AVERage <Average>
Activates automatic averaging, which means that the VNA may perform multiple cali-
bration sweeps and apply averaging to reduce trace noise. In contrast to manual aver-
aging (see [SENSe<Ch>:]AVERage[:STATe]) the number of calibration sweeps is
calculated automatically.
Parameters:
<Average> AUTO | MANual
Manual operation: See "Detector" on page 645
[SENSe:]CORRection:COLLect:CHANnels:ALL <Boolean>
Enables calibration of all channels in the active recall set. The command is valid for the
following calibration methods:
●
Manual system error correction
●
Automatic system error correction (cal unit)
●
Manual SMARterCal
●
Automatic SMARTerCal (cal unit)
A scalar power calibration is not affected.
Parameters:
<Boolean> ON – calibrate all channels.
OFF – calibrate the active channel only.
Example: See [SENSe<Ch>:]CORRection:COLLect:SAVE:
SELected[:DUMMy]
[SENSe<Ch>:]CORRection:COLLect:CHANnels:CSETup <Boolean>
Same as [SENSe:]CORRection:COLLect:CSETup, but applies to channel <Ch>
only.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> OFF (0)
Individual sweep setups for all standards (faster for some config-
urations)
SCPI command reference