Command reference
R&S
®
ZNA
1529User Manual 1178.6462.02 ─ 29
Parameters:
<FixedFreq> Fixed stimulus and analyzer frequency.
*RST: 1 GHz
Default unit: Hz
Example:
FUNC "XTIMe:POW:A1"
Activate a time sweep and select the wave quantity a
1
as mea-
sured parameter for channel and trace no. 1.
FREQ:CW 100MHz
Set the CW frequency to 100 MHz.
Manual operation: See "CW Frequency" on page 524
7.3.15.3 SOURce:INDependent...
The SOURce:INDependent... configure internal RF and LO ports as channel-inde-
pendent CW sources.
SOURce:INDependent:CLO:FREQuency <Frequency>
If the converter LO Out port is configured as an independent CW source (SOURce:
INDependent:CLO[:STATe] ON), this command sets the frequency of the CW sig-
nal.
Parameters:
<Frequency> Frequency of the CW signal
*RST: 1 GHz
Default unit: Hz
Manual operation: See "Frequency / Power" on page 921
SOURce:INDependent:CLO:OFF <RF off>
If the converter LO Out port is configured as an independent CW source (SOURce:
INDependent:CLO[:STATe] ON, this command toggles the CW signal.
Parameters:
<RF off> ON (1)
The independent CW source is switched off.
OFF (0)
The independent CW source is switched on.
*RST: OFF (0)
Manual operation: See "Source RF Off" on page 921
SOURce:INDependent:CLO:POWer <Power>
If the converter LO Out port is configured as an independent CW source (SOURce:
INDependent:CLO[:STATe] ON), this command sets the power of the CW signal.
SCPI command reference