Command reference
R&S
®
ZNA
1547User Manual 1178.6462.02 ─ 29
Manual operation: See "Control Loop Parameters" on page 531
SOURce<Ch>:POWer<Pt>:ALC:RANGe <CRange>
Defines the maximum change of the source signal level due to the ALC for physical
port <Pt> in channel <Ch>.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<CRange> Control range [dB]
Range: 0 to 100
Increment: 0.1
*RST: 5
Manual operation: See "Control Range +/-" on page 528
SOURce<Ch>:POWer<Pt>:ALC:RECeiver <ALCReceiver>
Defines the receiver to be used for ALC at port <Pt> in channel <Ch>.
The target level is taken from the port configuration (SOURce<Ch>:POWer<Pt>:ALC:
TPOWer:DEFault on page 1549 ON) or can be set using SOURce<Ch>:POWer<Pt>:
ALC:TPOWer.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<ALCReceiver> 'A<n>'
Reference receiver at analyzer port <n>
'B<n>'
Measurement receiver at analyzer port <n>
*RST: 'A<Pt>'
Manual operation: See "ALC: Meas Port/ALC: Meas Receiver" on page 529
SOURce<Ch>:POWer<Pt>:ALC:SOFFset <StartOffset>
Increases or decreases the signal level before the ALC loop is started for port <Pt> in
channel <Ch>.
Suffix:
<Ch>
.
Channel number
<Pt> Physical port number
SCPI command reference