EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1557

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1557User Manual 1178.6462.02 ─ 29
After defining the leveling grid (using
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency and
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer commands), you
can start the leveling data acquisition using SOURce:POWer<PhyPt>:CORRection:
SLEVeling[:ACQuire].
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<PowValue> Start/stop power
Default unit: dB
Example: See SOURce:POWer<PhyPt>:CORRection:SLEVeling[:
ACQuire]
Options: R&S ZNA-K8
Manual operation: See "Start Power/Stop Power" on page 603
SOURce:POWer<PhyPt>:CORRection:SLEVeling:POWer:STEP
<Size>
Defines the (source) power step size of the leveling grid for physical port <PhyPt>.
After defining the leveling grid (using
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency and
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer commands), you
can start the leveling data acquisition using SOURce:POWer<PhyPt>:CORRection:
SLEVeling[:ACQuire].
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<Size> Default unit: dB
Example: See SOURce:POWer<PhyPt>:CORRection:SLEVeling[:
ACQuire]
Options: R&S ZNA-K8
Manual operation: See "Power Step Size" on page 603
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:STATe <Active>
Defines whether converter power control is used, if a leveling dataset is available
(SOURce:POWer<PhyPt>:CORRection:SLEVeling:STATe? on page 1554).
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<Active>
Options: R&S ZNA-K8