EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1561

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1561User Manual 1178.6462.02 ─ 29
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration.
Example: See SOURce:POWer:CORRection:TCOefficient[:STATe]
on page 1563
Usage: Setting only
Manual operation: See "Frequency / Transm. Coefficients: Insert, Delete, Delete
All" on page 634
SOURce:POWer:CORRection:TCOefficient:DELete<ListNo>[:DUMMy]
[<DeEmbedding Position>]
Deletes entry <ListNo> from the transmission coefficient list.
Suffix:
<Ch>
.
Calibrated channel number. This suffix is ignored; the transmis-
sion coefficient settings are valid for all channels.
<PhyPt> Calibrated port number. This suffix is ignored; the transmission
coefficient settings are valid for all sources.
<ListNo> List entry no.
Parameters:
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration.
Example:
See SOURce:POWer:CORRection:TCOefficient[:STATe]
on page 1563
Manual operation: See "Frequency / Transm. Coefficients: Insert, Delete, Delete
All" on page 634