Command reference
R&S
®
ZNA
1571User Manual 1178.6462.02 ─ 29
Parameters:
<Boolean> Enables (ON|1) or disables (OFF|0) fast source power calibra-
tion.
*RST: OFF|0
SOURce<Ch>:POWer:CORRection:GENerator<Gen>[:STATe] <Boolean>
Enables or disables the source power calibration for channel <Ch> and for an external
generator number <Gen>. The command is disabled unless a source power calibration
for the external generator has been performed (SOURce<Ch>:POWer:CORRection[:
ACQuire]). To enable or disable a source power calibration for an analyzer port use
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe.
Suffix:
<Ch>
.
Calibrated channel number
<Gen> Generator number
Parameters:
<Boolean> Enables (ON) or disables (OFF) the source power calibration for
port number <PhyPt>.
*RST: OFF
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation: See "Port Overview" on page 637
SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire]
Starts the source calibration
●
at physical port <PhyPt>
●
in channel <Ch>
●
at harmonic orders [SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer?
?
Then stores and applies the calibration data.
Use SOURce:POWer:CORRection:PMETer:ID to select the suitable power meter.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Port number
SCPI command reference