Command reference
R&S
®
ZNA
1579User Manual 1178.6462.02 ─ 29
ONCE
The correction is applied once, at the initial start of the measure-
ment.
PSWeep
The correction is applied per sweep. from one sweep to the
next.
Manual operation: See "Premeas Mode" on page 533
SOURce<Ch>:POWer<Pt>:CPRemeas:NREadings <Max>
Defines the maximum number of cal precal sweeps to be performed in "once mode"
(SOURce<Ch>:POWer:CPRemeas:MODE ONCE).
This setting is channel- and port-specific.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Max> Maximum number of pre cal sweeps
Manual operation: See "Max Iterations/Tolerance/Convergence (table area)"
on page 532
SOURce<Ch>:POWer<Pt>:CPRemeas:NTOLerance <Tolerance>
Defines the maximum tolerated deviation between measured and target power during
cal premeas.
This setting is channel- and port-specific.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Tolerance> Maximum tolerated deviation in dB
Manual operation: See "Max Iterations/Tolerance/Convergence (table area)"
on page 532
SOURce<Ch>:POWer<Pt>:CPRemeas:RECeiver <CPRReceiver>
Defines the received signal to be used for cal premeas.
This setting is channel- and port-specific.
Suffix:
<Ch>
.
Channel number
<Pt> Port number