GUI reference
R&S
®
ZNA
367User Manual 1178.6462.02 ─ 29
Define
Selects the harmonic to be measured.
A "Frequency out of range" warning is displayed the harmonic frequency to be mea-
sured exceeds the frequency range of the analyzer. The harmonic traces are truncated
at the maximum receiver frequency.
"Parameter"
The following types of results are provided.
●
"Harmonic": The signal received at the Nth harmonic of the stimu-
lus signal is measured. The result is displayed as a wave quantity
but can be modified and converted using the functions in the
"Trace" softtool.
●
"Relative": The Nth harmonic is divided by the 1st harmonic (fun-
damental). This result is also called the Nth harmonic distortion
factor.
●
"THD_F" and "THD_R": The fundamental and root mean square
total harmonic distortion
See Chapter 4.7.3.3, "Harmonics measurements", on page 270
"Harmonic"
The Harmonic order can be set in the range between 2 (2nd har-
monic) and 100.
"Receiving
Port"
The receive port measures the signal at n times the fundamental fre-
quency where n is the harmonic order.
"Source Port"
The source port provides the stimulus signal at the fundamental fre-
quency.
Remote command:
CALCulate<Ch>:PARameter:MEASure '<TraceName>',
'h<Harmonic>b<Receiving Port>(p<Source Port>)[R|THDF|THDR]'
CALCulate<Ch>:PARameter:SDEFine '<TraceName>',
'h<Harmonic>b<Receiving Port>(p<Source Port>)[R|THDF|THDR]'
...
Properties
See "Properties" on page 361.
5.2.5 Noise Figure tab
Noise figure measurements are available for both non-frequency converting and fre-
quency converting DUTs.
The noise figure measurement is provided with software option R&S ZNA-K30. If this
option is not installed, the "Noise Figure" tab is not available.
For background information, see Chapter 4.7.11, "Noise figure measurement",
on page 289.
Access: Trace – [Meas] > "Noise Figure"
Meas softtool