Receiving Port
For the receiving port, you can activate or deactivate the Reverse Coupler Configura-
tion and the Internal low noise preamplifier. The R&S ZNA immediately calculates the
resulting measurement time and displays it in the Result Settings section.
In addition, it is possible to specify an external attenuation at the receiving port to let
the "Quickset" logic choose the suitable source powers for calibration and measure-
ment.
Remote command:
[SENSe<Ch>:]NFIGure:QSET:ATTReceiver
Det Time Overview
Opens the "Detector Time Overview" dialog, which shows an overview of the resulting
detector times for each step of the calibration and measurement.
The detector times are calculated based on the VNA characteristics. They ensure that
the selected trace noise is achieved over the intended frequency range, with minimum
overall measurement time.
Recommended Measurement and Calibration Settings
This section gives an overview of the power settings that the Quickset function recom-
mends for an ideal calibration and measurement, without compression of the DUT or
the VNA receivers. Using "Apply" or "OK" these values are directly transferred into the
corresponding measurement and calibration settings.
Apply/OK
Applies the settings of the quick NF setup. OK also closes the dialog.
Remote command:
[SENSe<Ch>:]NFIGure:QSET[:EXEC]