GUI reference
R&S
®
ZNA
453User Manual 1178.6462.02 ─ 29
you can use the time gate settings to eliminate unwanted responses in your signal.
After switching back to the frequency domain, you will receive the frequency response
of your DUT without the contribution of the unwanted responses. The time gate is inde-
pendent of the frequency window used to filter the trace before transforming it to time
domain.
Time domain analysis requires option R&S ZNA-K2. If this option is not installed, the
Time Domain tab and "Time Gate" tabs are hidden.
Background information
Refer to Chapter 4.7.2.5, "Time gates", on page 252.
Time Gate
Enables or disables the time gate for the time domain and frequency domain traces.
"Gat" is displayed in the trace list while the time gate is active.
Traces softtool