5.11.1.2 Controls on the Start Cal tab
All controls on the "Start Cal" are buttons and each of them opens a dialog or dock
widget wizard. The buttons are grouped in three sections.
●
"Cal Active Channel" section
The main purpose of the buttons in this section is to calibrate the current channel.
– "Quick Start Calibration..." immediately starts the calibration of the current
channel, using its current calibration configuration.
Opens the calibration wizard that guides you through the calibration data acqui-
sition.
– "Configure/Start Calibration..." opens the calibration setup dialog, which allows
you to configure the calibration setup of the current channel, and to start the
calibration. You can also switch to another channel and configure/start this
channel's calibration (see Chapter 4.2.2.3, "Multi-channel setup dialog",
on page 139).
●
"Calibrate All" section
The main purpose of the buttons in this section is to perform complex multi-channel
calibrations in an efficient way.
– "Quick Start Calibrate All..." immediately starts the calibration, using a previ-
ously configured subset of the available calibration setups, or all available cali-
bration setups.
Opens the calibration wizard that guides you through the calibration data acqui-
sition.
– "Configure/Start Calibrate All..." opens the Calibrate All dialog that allows you
to configure calibration setups for an arbitrary set of channels and to enable/
disable a subset of the configured calibration setups. Finally you can acquire
calibration data for the enabled calibration setups in an efficient way.
Cal softtool