GUI reference
R&S
®
ZNA
700User Manual 1178.6462.02 ─ 29
Acq Time/Fixed
Defines the duration of the data acquisition. "Fixed" means that Autoset does not
change the configured value.
Remote command:
[SENSe<Ch>:]PULSe:ATIMe
Bandwidth/Fixed
Defines the resolution bandwidth.
Remote command:
[SENSe<Ch>:]BWIDth[:RESolution]
Autoset
Adjusts Acq Delay, ALC Acq Delay, and Acq Time according to the configured pulses,
unless their value is fixed.
ALC Optimize
Enables/configures ALC optimization.
"Auto" (default)
The firmware decides whether speed optimization is performed.
"On"
Optimization active
"Off"
Optimization inactive
Remote command:
SOURce<Ch>:POWer<Pt>:ALC:OPTimize
IF Filter
See "IF Filter (analog)" on page 537
Sync Meas to Pulse Gen.
Allows you to switch between synchronous and asynchronous measurement.
Remote command:
[SENSe<Ch>:]PULSe:SYNCron[:STATe]
5.13.2.3 VSE Connection Configuration dialog
The controls on the "VSE Connection Configuration" dialog allow you to set up and
establish a connection to a local or remote R&S VSE instance.
For background information, see Chapter 4.7.6.1, "Live pulse analysis with R&S VSE",
on page 280.
Access: Pulse Mod tab > "Configure VSE"
Mode softtool