EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - 5.14.5.7 SFD Advanced Settings

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
740User Manual 1178.6462.02 ─ 29
Remote command:
CALCulate:FMODel:ISD:SCALe:FREQuency
DC Extrapolation ← Calculations
This setting is only active, if a recent version of the ISD tool (from 2019-12 or later) is
used.
If unchecked (default), the touchstone data of the generated deembedding file reach
down to the start frequency of the instrument sweep. If checked, the ISD tool adds
(extrapolated) S-parameter data for f = 0.
Remote command:
CALCulate:FMODel:ISD:DCEXtrapolat
Small Fixture ← Calculations
Enables/disables the ISD tool's small fixture mode.
Remote command:
CALCulate:FMODel:ISD:SMALlfixture
Reset to Default
Resets the settings of the "ISD Advanced Settings" to their defaults.
Remote command:
CALCulate:FMODel:ISD:PRESet
5.14.5.7 SFD Advanced Settings
Advanced settings of the SFD tool. For the analyzer firmware, these settings are
global.
2x Thru Port Ordering
Tells the SFD tool about the port ordering of the 2x Thru test coupon
"Ports 1&3 on Left" (default): Odd ports are on the left and even ports are on the
right
"Ports 1&2 on Left" ports 1 to N are on the left and ports N+1 to 2·N are on the
right
Remote command:
CALCulate:FMODel:SFD:DIFFcfg
Offset Embed softtool