GUI reference
R&S
®
ZNA
769User Manual 1178.6462.02 ─ 29
Measure ← 1L Measurement/2L Measurement/3L Measurement
Measures the full set of S-parameters at the selected ports (full 2- or 4-port).
Remote command:
CALCulate:FMODel:DELT:M1L:MEASure
CALCulate:FMODel:DELT:M2L:MEASure
CALCulate:FMODel:DELT:M3L:MEASure
Ports table ← 1L Measurement/2L Measurement/3L Measurement
Use the checkboxes in the "Active" column to include the respective port into the mea-
surement.
If Method 1L is used, four ports have to be selected in the "1L Measurement" section.
Remote command:
CALCulate:FMODel:DELT<Ph_pt>:M1L[:STATe]
CALCulate:FMODel:DELT<Ph_pt>:M2L[:STATe]
CALCulate:FMODel:DELT<Ph_pt>:M3L[:STATe]
Length ← 1L Measurement/2L Measurement/3L Measurement
Specifies the length of the related line.
Remote command:
CALCulate:FMODel:DELT:M1L:LENGth
CALCulate:FMODel:DELT:M2L:LENGth
CALCulate:FMODel:DELT:M3L:LENGth
Run
Runs the AITT-DL tool with the collected measurement data and the configured Delta-L
settings.
The firmware enables this action when sufficient data for the required lines are availa-
ble (either loaded or measured).
If the AITT-DL tool experiences errors, the analyzer firmware displays additional infor-
mation (see <Deembedding Tool> Errors dialog).
Remote command:
CALCulate:FMODel:DELT:RUN
Clear All 1L Measurements/Clear Selected 1L Measurements
If Method 1L is selected in the "Delta-L Settings" dialog, the VNA firmware builds up a
cache of coupon measurements (Touchstone files) in the
DeltaL\1LMeasurementCache subfolder of the user data folder
C:\Users\Public\Documents\Rohde-Schwarz\ZNA.
"Clear All 1L Measurements" deletes this cache, "Clear Selected 1L Measurements"
displays a "Delete files" dialog that allows you to select the cache files to be deleted.
Remote command:
CALCulate:FMODel:DELT:M1L:CACHe:CLEar:ALL
CALCulate:FMODel:DELT:M1L:CACHe:CLEar:SELected
Offset Embed softtool